CAUTIONIt is essential that all instructions in this manual be followed precisely to ensure proper operation of the equipment.STXplus Signal Transmitt
CHAPTER 3: STAND-ALONE ANALOG CALIBRATION AND SET UPCHAPTER 3: STAND-ALONE ANALOG CALIBRATION AND SET UPINTRODUCTIONThis chapter contains the proced
CHAPTER 3: STAND-ALONE ANALOG CALIBRATION AND SET UPSETTING UP CURRENTOUTPUT OPERATING MODE(0-20mA, 4-20mA, 20-4mA, and20-0mA)This procedure sets up
CHAPTER 3: STAND-ALONE ANALOG CALIBRATION AND SET UPNOTEThe STXplus must be in gross mode prior to calibration.ANALOG CALIBRATIONAnalog calibration c
CHAPTER 3: STAND-ALONE ANALOG CALIBRATION AND SET UPNOTEIf you do not press a button within15 seconds the STXplus exits theprogramming mode. Once st
CHAPTER 3: STAND-ALONE ANALOG CALIBRATION AND SET UPThe accuracy of the calibration will improve with a greater quantity of material addedduring the
CHAPTER 3: STAND-ALONE ANALOG CALIBRATION AND SET UPNOTEIf you do not press a button within15 seconds the STXplus exits theprogramming mode. Once st
CHAPTER 3: STAND-ALONE ANALOG CALIBRATION AND SET UPthe average. This results in fewer variations in output. For example, fewer variations in a di
CHAPTER 3: STAND-ALONE ANALOG CALIBRATION AND SET UPFigure 3-4. DSP Filter Examples.TAREThe tare feature allows you to tare or “zero” a vessel. Tar
CHAPTER 3: STAND-ALONE ANALOG CALIBRATION AND SET UPTable 3-2. 3-9
APPENDIX A: PRODUCT SPECIFICATIONSAPPENDIX A: PRODUCT SPECIFICATIONSANALOG INPUTResolution. Selectable 16 bit (1 part in 65,536) to 23 bit (1 part
NOTICEThe content of this document is the intellectual property of Kistler-Morse®. Any reproductionor translation of this document wihtout the writte
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSAPPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSINTRODUCTIONThis appendix contains the serial
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSB-2
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSB-3
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSTable B-2. Denitions of Characters in Command Strings.CHECKSUM CALCULATIONThe 1-byte checksum is
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSResponseThe response to a request for gross engineering units is:A(-)dddddddssrIftheweightis+6
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSRequest from Master Response from STXplus>aau1ssr A(-)dddddddssr Format >01u208r
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSRequest from Master Response from STXplus>aaP1???ssr Ar Format >01P1kgsC6r
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDS‘o’ Command — Default Calibration Constants OnlyThis command is sent by the master to request the
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDS‘Z’ Command — Perform Zero CalibrationThis command is sent by the master to request the STXplus ca
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSRequest from Master Response from STXplus>aan1ssr A000000dssr Format >01n100r
TABLE OF CONTENTSCHAPTER 1: INTRODUCTION 1-1Introduction 1-1Physical Description 1-1Functional Description 1-1 Analog Mode 1-1 Filters 1-1 L
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDS‘w1’ Command — Write Digital Delta (Scale Factor) CountsThis command is sent by the master to set
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSwould only be in effect when in digital mode. This would be the A/D counts required in order for
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSwhich would only be in effect when in digital mode. This would be the A/D counts required in orde
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSFor this example, the master requests the digital zero counts for an STXplus at address 01 which r
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSRequest from Master Response from STXplus>aaw8(-)fffffffssr Ar Format >01w814865.46
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSRequest from Master Response from STXplus>aaRCssr A(-)fffffffssr Format >01RCF6r
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSthe master requests the averaging value for an STXplus at address 01 which returns 34 and a checks
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSRequest from Master Response from STXplus>aaRXssr A0000dddssr Format >01RX0Br
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDS‘wZ’ Command — Write DSP Filter QualifyThis command is sent by the master to set the DSP Filter qu
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSRequest from Master Response from STXplus>aa[W200000ddssr Ar Format >01[W21578973r
System Set Up 3-6 Descriptions 3-6 Current Output Range 3-6 Current Output Tracking 3-6 Current Output Fail-safe 3-6 LO Span and HI Span 3-
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSexample, the master sets the current output high weight for an STXplus at address 01 assuming a fo
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDS‘bH’ Command — Write Current Output Fail-safe ModeThis command is sent by the master to set the cu
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDS‘bG’ Command — Write Current Output Tracking ModeThis command is sent by the master to set the cur
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDS‘PI’ Command — Write Setpoint Deadband ValueThis command is sent by the master to set the setpoint
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDS‘P9’ Command — Write Setpoint High/Low ModeThis command is sent by the master to set the setpoint
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDS‘wg’ Command — Write Setpoint StateThis command is sent by the master to set the setpoint state of
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDS‘PH’ Command — Write Setpoint ValueThis command is sent by the master to set the setpoint value of
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDS‘PT’ Command — Write Corrected/Out Linearization ValueThis command is sent by the master to set th
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSRequest from Master Response from STXplus>aaK2ssr A000000dssr Format >01K2DEr
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSRequest from Master Response from STXplus>aaL4000000dssr Ar Format >01L4112r
CHAPTER 1: INTRODUCTIONCHAPTER 1: INTRODUCTIONINTRODUCTIONThe STXplus Signal Transmitter, shown above, is a 4-channel signal processor that receives
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSRequest from Master Response from STXplus>aaK7ssr A000000dssr Format >01K7E3r
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDS Request from Master Response from STXplus>aaKAssr A000000dssr Format >01KAE
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDSRequest from Master Response from STXplus>aaLD0000dddssr Ar Format >01LD1254r
APPENDIX B: KISTLER-MORSE® PROTOCOL SERIAL COMMANDS‘g2’ Command — Write Serial Port 1 ProtocolThis command is sent by the master to set the serial po
APPENDIX C: MODBUS PROTOCOL SERIAL COMMANDSAPPENDIX C: MODBUS PROTOCOL SERIAL COMMANDSINTRODUCTIONThis appendix contains the serial commands and Mod
APPENDIX C: MODBUS PROTOCOL SERIAL COMMANDSTheCRCisstartedbyrstpreloadinga16-bitregistertoall1’s.Thenaprocessbeginsofapplying succe
APPENDIX C: MODBUS PROTOCOL SERIAL COMMANDSMODBUS FUNCTION CODES03 (0x03) Read Holding RegistersThis function code is used to read the contents of a
APPENDIX C: MODBUS PROTOCOL SERIAL COMMANDSErrorError code: (1 Byte) 0x90Exception code: (1 Byte) 01, 02, 03, or 045 (0x05) Write Single CoilThis fun
APPENDIX C: MODBUS PROTOCOL SERIAL COMMANDSMODBUS EXCEPTION RESPONSESWhen a request is sent, a normal response is expected. One of four possible even
APPENDIX C: MODBUS PROTOCOL SERIAL COMMANDSCOMMANDSTare RequestUsed to force a tare on the vessel.The response is an echo of the query after the Acti
CHAPTER 2: HARDWARE INSTALLATIONCHAPTER 2: HARDWARE INSTALLATIONGENERAL INFORMATIONThis chapter provides instructions on how to install and wire the
APPENDIX C: MODBUS PROTOCOL SERIAL COMMANDSRead a RegisterNOTEAddress xxxx must be a valid word address in the device or the device will respond with
APPENDIX C: MODBUS PROTOCOL SERIAL COMMANDSTable C-1. Modbus Address Map.C-8
APPENDIX C: MODBUS PROTOCOL SERIAL COMMANDSC-9Table C-1. Modbus Address Map. (continued)
APPENDIX C: MODBUS PROTOCOL SERIAL COMMANDSTable C-1. Modbus Address Map. (continued)C-10
APPENDIX C: MODBUS PROTOCOL SERIAL COMMANDSC-11Table C-1. Modbus Address Map. (continued)
APPENDIX C: MODBUS PROTOCOL SERIAL COMMANDSTable C-1. Modbus Address Map. (continued)C-12
APPENDIX D: TECHNICAL DRAWINGSAPPENDIX D: TECHNICAL DRAWINGSThis appendix contains the following technical drawings for the STXplus signal processor
APPENDIX D: TECHNICAL DRAWINGSD-2
APPENDIX D: TECHNICAL DRAWINGSD-3
APPENDIX D: TECHNICAL DRAWINGSD-4
CHAPTER 2: HARDWARE INSTALLATIONmost convenient to route the conduit orcables. Hole location is critical for proper PCB installation. Check clearan
APPENDIX D: TECHNICAL DRAWINGSD-5
APPENDIX D: TECHNICAL DRAWINGSD-6
APPENDIX D: TECHNICAL DRAWINGSD-7
APPENDIX D: TECHNICAL DRAWINGSD-8REVISIONS REV Description ECN Checked Approved Date A Initial Release 10-229 5-19-10Approvals Date ventureM
NOTES
All rights reserved.All data subject to change without notice.97-1180-01 Revision GJan 2012150 Venture BoulevardSpartanburg, SC 29306800.426.9010[T
CHAPTER 2: HARDWARE INSTALLATIONQUICK REFERENCEDefaultHold the UP, DOWN, and ENTER buttons while pressing the SW2 reset button.Changing the Modbus or
CHAPTER 2: HARDWARE INSTALLATIONEnabling the Modbus 5.11kΩ Pull-UpResistor and 12kΩ End-Of-Line ResistorThese jumpers enable and disable thepull-up r
Commentaires sur ces manuels